PRODUCTS

PLATINUM THIN FILM RTD TEMPERATURE SENSOR ELEMENTS

CRZ2005R-100

FEATURE

PLATINUM THIN FILM RTD TEMPERATURE SENSOR ELEMENT: Pt100

  1. All of elements are provided after inspected the actual resistance value at 0 . That is a great advantage for making temperature sensors if customers need similar or the same resistance value.
  2. This size of thin film RTD temperature sensor element is widely used in worldwide temperature metrology industry.
  3. The thin film RTD temperature sensor element achieved low cost due to the mass production system.
  4. Outstanding vibration and shock resistibility is improved comparing with wire wound types by circuit-sputter technology.
  5. The stability is excellent at even high temperature due to the construction.
  6. 10 pieces of thin film RTD temperature sensor elements are in a bag. (20 pieces for class 2B)
RoHS compliantマーク

CRZ2005R-100

PRODUCT CODE CRZ2005R-100
CLASS F0.1(1/3B), F0.15(A), F0.3(B), F0.6(2B)
OPERATING TEMP. RANGE F0.1(1/3B) -20 to 150℃
(-4 to+302 F)
F0.15(A) -40 to 300 ℃
(-40 to+572 F)
F0.3(B) -70 to 400℃
(-94 to+752 F)
F0.6(2B) -70 to 400℃
(-94 to+752 F)
DIMENSION(mm) (WxLxH) 2.0×5.0x1.1
THE NUMBER OF ELEMENT SINGLE
RESISTANCE VALUE Pt 100
MEASUREMENT CURRENT LESS THAN 1 mA
LEAD WIRE’S MATERIAL Au-Plated Nickel
LEAD WIRE’S DIMENSION (mm)
(DxL)
φ0.2×11
TEMPERATURE COEFFICIENT RESISTANCE (TCR) 0.003851
STABILITY 200℃, 1000 hour △R0 < ±0.02%
400℃, 1000 hour △R0 < ±0.04%
RESPOSE TIME
(90% RESPONSE)
AIR WATER
V=1.0 m/s V=3.0 m/s
16 11 0.3
SELF HEATING Condition Self-Heating(℃)
0.5mA 1mA (2mA)*
Still Air without MgO 0.08 0.47
With MgO Powder 0 0.07

*2mA for Pt100 is out of standard

TECHNICAL DATA

RESPONSE TIME

PLATINUM THIN FILM RTD TEMPERATURE SENSOR ELEMENT

Response time to 63.2% (time constant) and 90% of temperature shift in stir water and still air when temperature shift from T1 to T2 is 100% is as follow.

TYPE SIZE STIR WATER STILL AIR
63.2% 90% 63.2% 90%
PLATINUM THIN FILM RTD TEMPERATURE SENSOR ELEMENT 1632 0.3 0.7 4.3 9.9
2005 0.4 0.9 4.8 11.1
TEST CONDITION
TYPE OF ELEMENT THIN FILM RTD TEMPERATURE SENSOR ELEMENT
TEST TEMP. T1:LOW TEMP. → T2:HIGH TEMP.
T2:HIGH TEMP. → T1:LOW TEMP.
TEST METHOD The product’s measuring point is set in room temperature. Then, the response time is measured at time constant and 90% of response time after it is put in stir water and still air that have enough temperature difference with room temperature, instantly.
Response time is an average of T1 → T2 and T2 → T1.
STANDARD JIS C1604 RTD
Notification
This data is measured in a specified environment. By the actual measuring environment, and condition, the result will be different.

CONTACT US

ABOUT "CRZ2005R-100"

OTHER "PLATINUM THIN FILM RTD TEMPERATURE SENSOR ELEMENTS"

RTD element for thermometer

PLATINUM THIN FILM RTD TEMPERATURE SENSOR ELEMENTS

RoHS compliantマーク

CRZ1632R-100

RoHS compliantマーク

CRZ2005R-500

RoHS compliantマーク

CRZ2005R-1000

RoHS compliantマーク

CRZ1636RE-1000

RoHS compliantマーク

CRZ2005RE-1000

RoHS compliantマーク

CHARACTERISTICS OF RTD ELEMENTS

Contact Us

  • INQUIRY Mail Form