TECHNICAL DATA – RESPONSE TIME
FEP MOLD TYPE AND EXPOSED TYPES INSULATED THERMOCOUPLE
Response time to 63.2% (time constant) and 90% of temperature shift in stir water and still air when temperature shift from T1 to T2 is 100% is as follow.
OD(o) | STIR WATER | |||
---|---|---|---|---|
FEP MOLD 63.2% | EXPOSED 63.2% | FEP MOLD 90% | EXPOSED 90% | |
0.05 |
0.5 |
<0.01 |
1.3 |
0.01 |
0.32 |
2.9 |
0.15 |
6.6 |
0.35 |
0.64 |
4.9 |
0.27 |
11 |
0.63 |
/sec
OD(o) | STILL AIR | |||
---|---|---|---|---|
FEP MOLD 63.2% | EXPOSED 63.2% | FEP MOLD 90% | EXPOSED 90% | |
0.05 |
5.6 |
0.5 |
13 |
1.2 |
0.32 |
62 |
6.1 |
74 |
14.1 |
0.64 |
47 |
11.6 |
109 |
26.7 |
/sec
TEST CONDITION
TYPE OF ELEMENT WIRE
K FEP MOLD TYPE INSULATED THERMOCOUPLE
K EXPOSED TYPE INSULATED THERMOCOUPLE
TEST TEMP.
T1 : LOW TEMP.→T2 : HIGH TEMP.
T2 : HIGH TEMP.→T1 : LOW TEMP.
ROOM TEMP.→BOILED WATER
ROOM TEMP.→O
ROOM TEMP.→100 STILL AIR
100 STILL AIR→ROOM TEMP.
TEST METHOD
The product’s measuring point is set in room temperature. Then, the response time is measured at time constant and 90% of response time after it is put in stir water and still air that have enough temperature difference with room temperature, instantly.
Response time is an average of T1→T2 and T2→T1.
STANDARD
JIS C1602 THERMOCOUPLE
Notification
This data is measured in a specified environment.
By the actual measuring environment, and condition, the result will be different.
By the actual measuring environment, and condition, the result will be different.